The Water Flow
Laboratory has specialized test rigs for the precise calibration
of flowmeters, testing of valves, pipe fittings and other flow
products. The whole system is designed to handle a maximum flowrate
of 5000m3/h. Pipeline size upto 900mm. The gravimetric static
weighing method as per ISO 4185 is adopted here for the measurement
of flow rate.The two weighing tanks are of capacities 2T &
20T. Diverter systems are used to collect water in the weighing
tank, for an accurately measured period of time. The mass measurement
is by load cells. The relative density together with measurements
of water temperature at the moment of each diversion enables the
mass readings after correction for air buoyancy, to be converted
into volume. These volumes are divided by the respective diversion
times to obtain accurate measurement of flow rate.